Model 4001A, 4003A
Sweep Function Generator
INSTRUCTION MANUAL
Przeglądanie stron 0
1 2 3 4 5 6 ... 18 19

Podsumowanie treści

Strona 1 - Sweep Function Generator

Model 4001A, 4003A Sweep Function Generator INSTRUCTION MANUAL

Strona 2

10 Figure 1 – Front Panel Controls (4001A) Figure 2 - Front Panel Controls (4003A)

Strona 3 - Section 1

11 10. FUNCTION SELECTOR SWITCH A rotary switch for waveform selection. 11. FREQUENCY RANGE SELECTOR SWITCH A rotary switch to s

Strona 4

12 Figure 3 - Rear Panel

Strona 5 - Section 2

13 Section 5 Operating Instructions Before applying power to unit, make sure that input voltage setting is correct and the ventilation holes are

Strona 6 - Section 3

14 4. Display will show the frequency and Hz / kHz LED’s will light depending on the frequency. 5.4 VOLTAGE CONTROLLED FREQUENCY

Strona 7

15 Section 6 Maintenance 6.1 PREVENTIVE MAINTENANCE Please follow the following preventive steps to ensure the proper operation of your instr

Strona 8

16 Because of the specialized skills and test equipment required for instrument repair and calibration, many customers prefer to rely

Strona 9 - Section 4

17 Section 8 SERVICE INFORMATION Warranty Service: Please return the product in the original packaging with proof of purchase to the address

Strona 10

18 This warranty gives you specific rights and you may have other rights, which vary from state-to-state. B&K Precision Corp. 22820 Savi Ranch

Strona 11

19 Printed in China v040814

Strona 12 - Figure 3 - Rear Panel

2 Table of Contents Section 1!SAFETY GUIDELINES ... 3!COMPLIANCE STA

Strona 13 - Section 5

3 Section 1 Safety Guidelines WARNING Normal use of test equipment exposes you to a certain amount of danger from electrical shock because testing m

Strona 14

4 Compliance Statements Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with

Strona 15 - Section 6

5 Section 2 Introduction Models 4001A and 4003A are 4 MHz sweep function generators. Model 4003A has an added feature of a 20 MHz digital count

Strona 16 - Section 7

6 Section 3 Technical Specifications 3.1 FREQUENCY CHARACTERISTICS Waveforms: Sine, Square, Triangle, Ramp, ± Pulse Range: 0.5 Hz to 4 MHz i

Strona 17 - Section 8

7 3.6 SQUARE WAVE Symmetry: 50 % (positive half) to 50 % (negative half) < 2%, 1 Hz to 100 kHz Rise Time: < 90 ns, (20 Vpp, no

Strona 18

8 3.13 FREQUENCY COUNTER (model 4003A only) Display: 5-digit, auto range Display Unit: Hz/kHz auto range Resolution: 0.01 Hz (max.) Ac

Strona 19

9 Section 4 Controls & Indicators 4.1 FRONT PANEL (Refer to Figure 1 & Figure 2) 1. POWER SWITCH Pushing the switch “ON” will l