PROGRAMMING MANUAL Function/Arbitrary Waveform Generator MODEL: 4050 Series (4052, 4053, 4054, 4055)

Strona 2

6 1.4. Output Command DESCRIPTION Enable or disable the output of the [Output] connector at the front panel corresponding to the channel. The q

Strona 3 - Table of Contents

7 1.5. Basic Wave Command DESCRIPTION Set or get basic wave parameters. COMMAND SYNTAX <channel>:BaSicWaVe <parameter>

Strona 4

8 Maximal is 6V. Channel two minimum value 0.004V, Maximal is 20V. } <offset>:= { Default unit is "V". maximal value depends

Strona 5 - 1.1.4. Command Notation

9 <channel>:={C1, C2} <parameter>:= {a parameter from the table below. } Parameters Value Describe STATE <s

Strona 6

10 PWM,FRQ <pwm frequency> PWM frequency. Only carry wave is PULSE wave. PWM,DEVI <p wm devi > Duty cycle deviation. Only carry wave

Strona 7

11 <deep>:= {0% to 120%} <fm frequency>:= {0.002Hz to 20000Hz} <fm frequency offset> :={0 to basic wave fr

Strona 8 - 1.3.2 OPC

12 AMP,FRQ,1000HZ,AMP,4V,OFST,0V,SYM,50 EXAMPLE 3 Reads channel one modulate wave parameters that STATE is OFF. C1:MDWV?

Strona 9 - 1.3.3 IDN

13 1.7. Sweep Wave Command DESCRIPTION Set or get sweep wave parameters. COMMAND SYNTAX <channel>:SWeepWaVe <parameter>

Strona 10 - 1.4. Output Command

14 where: <state>:= {ON|OFF} <time>:= {0.001S to 500S} <stop frequency> :={ the same with basic wave fre

Strona 11 - 1.5. Basic Wave Command

15 EXAMPLE 4 Reads channel two modulate wave parameters that STATE is OFF. C2:SWWV? Return: C2:SWWV STATE,OFF 1.8. Bur

Strona 13

16 EDGE <edge> When carrier wave is NOISE wave, you can’t set it. When NCYC is selected and TRSR is set to EXT, you can set it. TIME <cir

Strona 14

17 <offset>:={ Default unit is "V". the maximal value depends on the version.} <duty>:={ If wave type is Square, range is from

Strona 15

18 1.9. Parameter Copy Command DESCRIPTION copy channel data. COMMAND SYNTAX PAraCoPy <destinat channle>, <src channle>

Strona 16

19 ne 11 ExpRise 22 Lorentz 36 Kaiser 47 Acos 12 LogFall 23 Gauspuls 37 BlackMan 48 Atan QUERY SYNTAX <channel>:ARbWaVe?

Strona 17 - 1.7. Sweep Wave Command

20 1.11. Phase Command DESCRIPTION Set or get phase parameters. COMMAND SYNTAX INVerT <parameter> <parameter>

Strona 18

21 1.13. Configuration Command DESCRIPTION Changes system load data of power on. COMMAND SYNTAX Sys_CFG<parameter>

Strona 19 - 1.8. Burst Wave Command

22 QUERY SYNTAX SCreen_SaVe? RESPONSE FORMAT SCreen_SaVe <parameter> EXAMPLE 1 Set screen save time 5 minutes. SC

Strona 20

23 TRG <trig leve l> Input signal trig level. PW <positive width> Input signal positive width. NW <negative width> Input signal n

Strona 21

24 1.18. Store list command DESCRIPTION This command is used to read the device wave data name. If the store unit is empty, the command will

Strona 22 - 1.10. Arbitrary Wave Command

25 1.19. Get arbitrary wave data command DESCRIPTION This command changes the user defined memory unit arbitrary wave data. COMMAND SYNTAX

Strona 23

Table of Contents 1.1. About Commands & Queries ______________________________________ 1 1.2. Table of Commands & Queries ________________

Strona 24 - 1.11. Phase Command

26 \x01\x0e\x01\x11\x01\x14\x01\x17\x01\x1a\x01\x1d\x01 \x01$\x01'\x01*\x01-\x010\x013\x016\x01:\x01=\x01@\x01C\x01F\x01I\x01L\x01P\x01S\x01V

Strona 25 - 1.15. Screen Save Command

27 1.20. Virtual key command DESCRIPTION This sends a virtual key command to the device. The keys are representations of the front panel but

Strona 27

V042413 1.21. I n dex *IDN *IDN *OPC *OPC A ARWV ARBWAVE B BSWV BASIC_WAVE BTWV BURSTWAVE BUZZ BUZZER C CHCP CHANNEL_COPY CHDR COMM_HEADER D DCW

Strona 29

1 1.1. About Commands & Queries This section lists and describes the remote control commands and queries recognized by the instrument. All comm

Strona 30

2 placeholders, of which there are two types: the header path and the data parameter of a command. := A colon followed by an equals sign sepa

Strona 31 - 1.20. Virtual key command

3 1.2. Table of Commands & Queries Short Long Form Subsystem What Command/Query dose *IDN *IDN SYSTEM Get identification from device . *OPC *O

Strona 32

4 1.3. IEEE 488.2 Common Command Introduction IEEE standard defines the common commands used for querying the basic information of the instrume

Strona 33 - 1.21. I n dex

5 1.3.3 IDN DESCRIPTION The *IDN? Query causes the instrument to identify itself. The response comprises manufacturer, mode l number, serial num