4
2. Specifications
2.1 Performance
Frequency section
2650A, 2652A, 2651 2658A
―――
GHz
resolution
Allows rotary encoder, numeric key and function key
within ± (30+20T) kHz±1 dot
@frequency span: 200 kHz to 10 MHz,
RBW 3 kHz, 23±5°C
within ± (60+300T) kHz±1dot
@frequency span: 20 MHz to 3.3 GHz,
RBW 100 kHz、23±5°C
T: sweep time (s)
within ±(30+20T) kHz ± 1 dot
@frequency span: 200 kHz to 10 MHz,
RBW 3 kHz, 23±5°C
within ± (60+300T) kHz±1dot
@frequency span: 20 MHz to 8.5 GHz,
RBW 100 kHz, 23±5°C
T: sweep time (s)
frequency
error
within ±4 kHz @ 3 kHz, 10 kHz, 30 kHz
within 20% of RBW @ RBW: 100 kHz, 300 kHz
within 10% of RBW @ RBW: 1 MHz, 3 MHz
200 kHz to 2 GHz (1-2-5 step) and
3.3 GHz (full span)
200 kHz to 5 GHz (1-2-5 step) and
8.5 GHz (full span)
@ sweep time of 0.3s, 23±5 °C
# of points per trace
501dots on LCD screen, 1001dots readout via PC
(501 dots are visible on the display, 1001 dots of trace data are captured internally and can be
transferred to a PC via USB device interface.)
3 dB bandwidth (6 dB for 2651 @ 9 kHz, 120 kHz)
3 kHz to 3 MHz (1-3 step) and AUTO
(2651: 3 kHz, 9 kHz, 30 kHz, 120 kHz, 300 kHz, 1 MHz, 3 MHz)
1:12 (typical, 3 dB : 60 dB)
100 Hz to 1 MHz (1-3 step) and AUTO
@100 kHz offset, RBW : 3 kHz, VBW : 100 Hz, sweep time: 1 s
less than -40 dBc @
100 MHz
T:掃引時間