Bk-precision 5492BGPIB - Manual Instrukcja Użytkownika Strona 99

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 111
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 98
SCPI Command Reference
99
:AUTO <b>
Command Syntax:
:TRIGger:DELay:AUTO <b>
Command parameter:
<b> = 1 or ON Enable auto delay
0 or OFF Disable auto delay
Query:
:AUTO? Query delay state
Return String:
1 or 0
Description:
This command is used for enable or disable trigger auto delay. If disabled, the trigger will delay
based on the delay time configured with “:TRIGger:DELay” command.
6.3.8 Common Commands
Common commands can be used in instruments that are SCPI-compliant. The common commands
available for the 5492B are described below:
*RST
Command Syntax:
*RST
Description:
Resets the instrument.
*TRG
Command Syntax:
*TRG
Description:
Sends a bug trigger to the instrument. Use this if trigger source is set to BUS.
*IDN?
Query Syntax:
*IDN?
Query return:
<product>,<version>,<sn number>
Example: 5492B Digital Multimeter, Ver1.0.00.00.01,123A45678
Description: Query the identification of the instrument.
Przeglądanie stron 98
1 2 ... 94 95 96 97 98 99 100 101 102 103 104 ... 110 111

Komentarze do niniejszej Instrukcji