internal trigger source.
Example: CALC:FUNC AVER
Note: The trigger source is changed to
immediate when the AVERage
function is enabled.
User can combine TOL/NULL,
LIM/NULL, AVER/NULL or
NULL/AVER functions, but the other
combinations are void.
CALCulate:FUNCtion?
Description: Query the function of the
calculation.
Response: Return <Literal> format string:
AVER, TOL, LIM, NULL or NONE.
For combined modes, string can be:
TOL,NULL; LIM,NULL; AVER,NULL
Example: Returning AVER, NULL means the
record function and the relative
function being enabled.