QUERY SYNTAX WaveForm_SetUp?
Note 1: After power-on, SP is set to 4, NP is set
to 1000, and FP is set to 0.
Note 2: Parameters are grouped in pairs. The
first of the pair names the variable to be
modified, whilst the second gives the
new value to be assigned. Pairs may be
given in any order and may be restricted
to those variables to be changed.
RESPONSE FORMAT WaveForm_SetUp SP, <sparsing>, NP,
<number>, FP, <point>
EXAMPLE The following command specifies that every 3rd
data point (SP=3) starting at address 200 should
be transferred:
Command message:
WFSU SP, 3, FP, 200
RELATED COMMANDS WAVEFORM