DESCRIPTION The AUTO_CALIBRATE command is used to enable
or disable the quick calibration of the instrument.
The quick calibration may be disabled by issuing the
command ACAL OFF. Whenever it is convenient, a
*CAL? Query may be issued to fully calibrate the
oscilloscope.
The response to the AUTO_CALIBRATE?
Query indicates whether quick -calibration is enabled.
The command is only used in the CFL series
instrument.
COMMAND SYNTAX Auto_CALibrate <state>
<state> : = {ON, OFF}
QUERY SYNTAX Auto_CALibrate?
RESPONSE FORMAT Auto_CALibrate <state>
EXAMPLE The following instruction disables quick-calibration:
Command message:
ACAL OFF
RELATED COMMANDS *CAL?
MISCELLANEOUS
AUTO_CALIBRATE, ACAL